Max. DAC output voltage | 150 kVpeak / 106 kVrms | DAC frequency range | 20 Hz … 500 Hz | Capacitance range | 0.025 µF … 13 µF | HV charging current | 10 mA | PD measuring range | 1 pC … 100 nC | PD level detection | acc. to IEC 60270 | Bandwidth for PD-localisation | 150 kHz … 45 MHz | Dissipation factor δ | 0.1 % … 10 % | Power supply | 115 / 230 V, 50 / 60 Hz | Operating temperature | -10 °C … +40 °C | Weight | app. 300 kg | | | Components | | Unit 1 | HV source and LTT switch | Unit 2 (3 coils &basis) | Ø: 650 mm, H: 1400 mm | Unit 3 | HV divider and DAQ | Unit 4 | Notebook app. 2 kg | |
 |
| |
Dielectric Diagnosis and Partial Discharge Measuring Printversion Top
|
|
| |
|
|
|
|